Effect of fetal calf serum on the corrosion behaviour of magnesium alloys

Research output: Journal contributionsJournal articlesResearchpeer-review

Standard

Effect of fetal calf serum on the corrosion behaviour of magnesium alloys. / Hornberger, H.; Witte, F.; Hort, N. et al.

In: Materials Science and Engineering: B, Vol. 176, No. 20, 15.12.2011, p. 1746-1755.

Research output: Journal contributionsJournal articlesResearchpeer-review

Harvard

APA

Vancouver

Hornberger H, Witte F, Hort N, Mueller WD. Effect of fetal calf serum on the corrosion behaviour of magnesium alloys. Materials Science and Engineering: B. 2011 Dec 15;176(20):1746-1755. doi: 10.1016/j.mseb.2011.07.018

Bibtex

@article{77f66854b4754ec594a401c240f68025,
title = "Effect of fetal calf serum on the corrosion behaviour of magnesium alloys",
abstract = "The corrosion behaviour of WE43 magnesium alloys using the mini cell system was studied. Voltammetry and impedance spectroscopy were applied to study on the one hand the effect of microstructure of the working electrode and on the other hand the effect of proteins in the electrolyte. Two types of alloy samples were produced (i) by permanent mould casting and (ii) by gas atomization followed by extrusion. The results showed that the microstructure was strongly influenced by the production process. The extruded samples showed an improved homogeneity of phase distribution compared with cast samples as it was aimed for. Due to increased homogeneity it was expected to find higher corrosion resistance. However, the electrochemical results are contradictory and suggest an additional phase in the extruded microstructure. Using energy dispersive X-ray spectroscopy (EDX) the secondary magnesium rare earths (RE) phase of extruded samples showed differing composition than of cast samples as well as additional oxide phases. After the samples were electrochemically investigated in cell medium with and without fetal calf serum (FCS), an impact of FCS was detected in voltammetry due to the length of the polarisation curve. As the tip of the mini cell in contact with the working electrode is small, developing gases tend to spread on the working electrode and break the contact of liquid to the counter electrode; which results in disrupting the current flow. This effect was more pronounced when rising the voltage and was found reduced when using electrolytes with FCS. Impedance spectra were slightly deformed by FCS, seen as a kinetic effect but not as a basic differing corrosion reaction. The insight into the effects of FCS was provided by the mini cell system as this system enables to collect entire series of measurements. In contrast of two single measurements, those series reflected the slight difference caused by FCS. The focus of the electrochemical corrosion study was set on the first half hour of immersion.",
keywords = "Corrosion, Impedance, Magnesium, Polarization, Proteins, Reproducibility, Engineering",
author = "H. Hornberger and F. Witte and N. Hort and Mueller, {W. D.}",
year = "2011",
month = dec,
day = "15",
doi = "10.1016/j.mseb.2011.07.018",
language = "English",
volume = "176",
pages = "1746--1755",
journal = "Materials Science and Engineering: B",
issn = "0921-5107",
publisher = "Elsevier B.V.",
number = "20",

}

RIS

TY - JOUR

T1 - Effect of fetal calf serum on the corrosion behaviour of magnesium alloys

AU - Hornberger, H.

AU - Witte, F.

AU - Hort, N.

AU - Mueller, W. D.

PY - 2011/12/15

Y1 - 2011/12/15

N2 - The corrosion behaviour of WE43 magnesium alloys using the mini cell system was studied. Voltammetry and impedance spectroscopy were applied to study on the one hand the effect of microstructure of the working electrode and on the other hand the effect of proteins in the electrolyte. Two types of alloy samples were produced (i) by permanent mould casting and (ii) by gas atomization followed by extrusion. The results showed that the microstructure was strongly influenced by the production process. The extruded samples showed an improved homogeneity of phase distribution compared with cast samples as it was aimed for. Due to increased homogeneity it was expected to find higher corrosion resistance. However, the electrochemical results are contradictory and suggest an additional phase in the extruded microstructure. Using energy dispersive X-ray spectroscopy (EDX) the secondary magnesium rare earths (RE) phase of extruded samples showed differing composition than of cast samples as well as additional oxide phases. After the samples were electrochemically investigated in cell medium with and without fetal calf serum (FCS), an impact of FCS was detected in voltammetry due to the length of the polarisation curve. As the tip of the mini cell in contact with the working electrode is small, developing gases tend to spread on the working electrode and break the contact of liquid to the counter electrode; which results in disrupting the current flow. This effect was more pronounced when rising the voltage and was found reduced when using electrolytes with FCS. Impedance spectra were slightly deformed by FCS, seen as a kinetic effect but not as a basic differing corrosion reaction. The insight into the effects of FCS was provided by the mini cell system as this system enables to collect entire series of measurements. In contrast of two single measurements, those series reflected the slight difference caused by FCS. The focus of the electrochemical corrosion study was set on the first half hour of immersion.

AB - The corrosion behaviour of WE43 magnesium alloys using the mini cell system was studied. Voltammetry and impedance spectroscopy were applied to study on the one hand the effect of microstructure of the working electrode and on the other hand the effect of proteins in the electrolyte. Two types of alloy samples were produced (i) by permanent mould casting and (ii) by gas atomization followed by extrusion. The results showed that the microstructure was strongly influenced by the production process. The extruded samples showed an improved homogeneity of phase distribution compared with cast samples as it was aimed for. Due to increased homogeneity it was expected to find higher corrosion resistance. However, the electrochemical results are contradictory and suggest an additional phase in the extruded microstructure. Using energy dispersive X-ray spectroscopy (EDX) the secondary magnesium rare earths (RE) phase of extruded samples showed differing composition than of cast samples as well as additional oxide phases. After the samples were electrochemically investigated in cell medium with and without fetal calf serum (FCS), an impact of FCS was detected in voltammetry due to the length of the polarisation curve. As the tip of the mini cell in contact with the working electrode is small, developing gases tend to spread on the working electrode and break the contact of liquid to the counter electrode; which results in disrupting the current flow. This effect was more pronounced when rising the voltage and was found reduced when using electrolytes with FCS. Impedance spectra were slightly deformed by FCS, seen as a kinetic effect but not as a basic differing corrosion reaction. The insight into the effects of FCS was provided by the mini cell system as this system enables to collect entire series of measurements. In contrast of two single measurements, those series reflected the slight difference caused by FCS. The focus of the electrochemical corrosion study was set on the first half hour of immersion.

KW - Corrosion

KW - Impedance

KW - Magnesium

KW - Polarization

KW - Proteins

KW - Reproducibility

KW - Engineering

UR - http://www.scopus.com/inward/record.url?scp=81155162538&partnerID=8YFLogxK

U2 - 10.1016/j.mseb.2011.07.018

DO - 10.1016/j.mseb.2011.07.018

M3 - Journal articles

AN - SCOPUS:81155162538

VL - 176

SP - 1746

EP - 1755

JO - Materials Science and Engineering: B

JF - Materials Science and Engineering: B

SN - 0921-5107

IS - 20

ER -