A New and Rapid Method for Monitoring the New Oxazolidinone Antibiotic Linezolid in Serum and Urine by High Performance Liquid Chromatography-Integrated Sample Preparation

Research output: Journal contributionsJournal articlesResearchpeer-review

Authors

A sensitive and rapid HPLC-assay for determining the new oxazolidinone antibiotic linezolid in serum and urine is described. HPLC-integrated sample preparation permits the direct injection of serum and urine samples without any pre-treatment. The in-line extraction technique is realized by switching automatically from the extraction column to the analytical column. After the matrix has passed the extraction column the retained analyte will be quantitatively transferred to the analytical column where separation by isocratic HPLC will be performed. Linezolid is detected according to its absorption maximum at 260 nm. The quantification limits are estimated to be 0.3 and 0.5 μg/ml in serum and urine samples, respectively. The described procedure allows sample clean-up and determination of the antibiotic within 20 min, thereby facilitating drug-monitoring in clinical routine.

Original languageEnglish
JournalJournal of Chromatography B
Volume755
Issue number1-2
Pages (from-to)373-377
Number of pages5
DOIs
Publication statusPublished - 05.05.2001
Externally publishedYes

    Research areas

  • Linezolid, Oxazolidinones
  • Chemistry