General strategies to increase the repeatability in non-target screening by liquid chromatography-high resolution mass spectrometry
Publikation: Beiträge in Zeitschriften › Zeitschriftenaufsätze › Forschung › begutachtet
Standard
in: Analytica Chimica Acta, Jahrgang 935, 07.09.2016, S. 173-186.
Publikation: Beiträge in Zeitschriften › Zeitschriftenaufsätze › Forschung › begutachtet
Harvard
APA
Vancouver
Bibtex
}
RIS
TY - JOUR
T1 - General strategies to increase the repeatability in non-target screening by liquid chromatography-high resolution mass spectrometry
AU - Bader, Tobias
AU - Schulz, Wolfgang
AU - Kümmerer, Klaus
AU - Winzenbacher, Rudi
N1 - Copyright © 2016 Elsevier B.V. All rights reserved.
PY - 2016/9/7
Y1 - 2016/9/7
N2 - This article focuses on the data evaluation of non-target high-resolution LC-MS profiles of water samples. Taking into account multiple technical replicates, the difficulties in peak recognition and the related problems of false positive and false negative findings are systematically demonstrated. On the basis of a combinatorial approach, different models involving sophisticated workflows are evaluated, particularly with regard to the repeatability. In addition, the improvement resulting from data processing was systematically taken into consideration where the recovery of spiked standards emphasized that real peaks of interest were barely or not removed by the derived filter criteria. The comprehensive evaluation included different matrix types spiked with up to 263 analytical standards which were analyzed repeatedly leading to a total number of more than 250 injections that were incorporated in the assessment of different models of data processing. It was found that the analysis of multiple replicates is the key factor as, on the one hand, it provides the option of integrating valuable filters in order to minimize the false positive rate and, on the other hand, allows correcting partially false negative findings occurring during the peak recognition. The developed processing strategies including replicates clearly point to an enhanced data quality since both the repeatability as well as the peak recognition could be considerably improved. As proof of concept, four different matrix types, including a wastewater treatment plant (WWTP) effluent, were spiked with 130 isotopically labeled standards at different concentration levels. Despite the stringent filter criteria, at 100 ng L(-1) recovery rates of up to 93% were reached in the positive ionization mode. The proposed model, comprising three technical replicates, filters less than 5% and 2% of the standards recognized at 100 and 500 ng L(-1), respectively and thus indicates the general applicability of the presented strategies.
AB - This article focuses on the data evaluation of non-target high-resolution LC-MS profiles of water samples. Taking into account multiple technical replicates, the difficulties in peak recognition and the related problems of false positive and false negative findings are systematically demonstrated. On the basis of a combinatorial approach, different models involving sophisticated workflows are evaluated, particularly with regard to the repeatability. In addition, the improvement resulting from data processing was systematically taken into consideration where the recovery of spiked standards emphasized that real peaks of interest were barely or not removed by the derived filter criteria. The comprehensive evaluation included different matrix types spiked with up to 263 analytical standards which were analyzed repeatedly leading to a total number of more than 250 injections that were incorporated in the assessment of different models of data processing. It was found that the analysis of multiple replicates is the key factor as, on the one hand, it provides the option of integrating valuable filters in order to minimize the false positive rate and, on the other hand, allows correcting partially false negative findings occurring during the peak recognition. The developed processing strategies including replicates clearly point to an enhanced data quality since both the repeatability as well as the peak recognition could be considerably improved. As proof of concept, four different matrix types, including a wastewater treatment plant (WWTP) effluent, were spiked with 130 isotopically labeled standards at different concentration levels. Despite the stringent filter criteria, at 100 ng L(-1) recovery rates of up to 93% were reached in the positive ionization mode. The proposed model, comprising three technical replicates, filters less than 5% and 2% of the standards recognized at 100 and 500 ng L(-1), respectively and thus indicates the general applicability of the presented strategies.
KW - Chemistry
KW - Data processing
KW - Feature validation
KW - High-resolution mass spectrometry
KW - Non-target screening
KW - Peak detection
KW - Repeatability
UR - http://www.scopus.com/inward/record.url?scp=84977659373&partnerID=8YFLogxK
U2 - 10.1016/j.aca.2016.06.030
DO - 10.1016/j.aca.2016.06.030
M3 - Journal articles
C2 - 27543026
VL - 935
SP - 173
EP - 186
JO - Analytica Chimica Acta
JF - Analytica Chimica Acta
SN - 0003-2670
ER -